LSR II
Overview
The LSR II is a digital three-laser multi-colour analysis system (Blue 488 nm, Red 633 nm and Violet 405 nm) with the capacity of detecting up to 14 colours in addition to FSC and SSC.
Detailed information
| LASER | COLOUR | FILTER | MIRROR |
|---|---|---|---|
| Blue 488 nm | FITC, Alexa 488 | 530/30 | 505 LP |
| PE | 575/25 | 550LP | |
| PerCP, PerCP-CY5.5 | 695/40 | 685LP | |
| PE-Cy5, PE-Cy5.5 | 660/20 | 655LP | |
| PE-Texas Red | 610/20 | 600LP | |
| PE-Cy7 | 780/60 | 755LP | |
| Red 633 nm | APC, Alexa 647 | 660/20 | |
| APC-Cy7, APC-H7 | 780/60 | 755LP | |
| APC-Alexa 700 | 730/45 | 710LP | |
| APC-680 | 635/35 | 675LP | |
| Violet 405 nm | Pacific Blue, BV421 | 450/50 | |
| AmCyan, BV510 | 525/50 | 505LP | |
| BV605 | 610/20 | 595LP | |
| BV650 | 670/30 | 630LP | |
| BV711 | 710/50 | 690LP | |
| BV785 | 780/50 | 750LP |
- Please note – PE-Cy5, PE-Cy5.5, perCP and PerCP-Cy5.5 cannot be used together.
- PE-Cy5 and APC may be used together for molecules that are not mutually expressed by cells.
- Alexa -700 and Alexa- 680 cannot be used together.
LSR II has the capability for adjusting the flow rate for various applications. The three preset flow rates are: LO = 12 µl/min, MED = 35 μl/min, HI = 60 μl/min.
LSR II is equipped with FACSFLOW supply system fluidics cart that incorporate a 20 L FACSFLOW cubitainer and a 10 L waste collection container. The Fluidics sensors monitor the level of sheath fluid or waste and warn when sheath fluid is low or empty, or when the waste container is full.